39
Operations & Maintenance Manual: Cleaning & Disinfecting
COMPATIBILITY
Table 6 shows the disinfectant and cleaning products that have demonstrated compatibility with system
materials but have not been tested for biological efficacy. Solutions included in the cleaning, disinfection, and
sterilization procedures have demonstrated both compatibility and biological efficacy.
Results with compatible solutions may vary based on exposure periods and device handling. Ensure that you
adhere to an inspection schedule as described in Periodic Inspections on page56.
Availability of cleaning and disinfection products varies by country, and Verathon is unable to test products in
every market. For more information, please contact Verathon Customer Care or your local representative. For
contact information, visit verathon.com/support.
Table 6. Compatible Pre‑Cleaners, Cleaning Solutions, and Disinfectants
PRODUCT
CYCLES PER COMPONENT
AVL Video
Baton
Video
Baton 2.0
Smart Cable
Ecolab
®
MetalClean 1500
Ecolab
®
Enzymatic Detergent 1000 2000 1500
Ecolab
®
OptiPro™ Gel Instrument Pre‑Cleaner 1500
schülke
®
thermosept
®
ER cleaner 1500
Anios
®
Anioxyde
®
1000 high level disinfectant and cold sterilant 1500
neodisher
®
MediClean forte 1500
Anios
®
Aniosyme
®
XL3 enzymatic detergent 1500
STERIS
®
PRE‑Klenz™ Instrument Transport Gel 1000 2000 1500
Certol
®
ProEZFoam™ 500
Getinge
®
Renuzyme Foam Spray 250
After reviewing Table 5 and Table 6, complete the following procedures to clean, disinfect, or sterilize the
system components:
• Clean the GlideScope Video Monitor
• Clean the Power Adapter
• Clean the Cradle
• Clean the GlideScope Premium Cart
• Remove the Stat
• Clean & Disinfect the Video Baton
• Inspect the Video Baton
• Clean the Smart Cable
• Disinfect the Smart Cable
• Sterilize the Smart Cable (Optional)
• Clean & Disinfect the GlideScope Direct