7.4.2.5 Set Time-Based Recording Delay Command
Length
Command
Group
Command
Code
Card Mask Channel
Mask
Parameters
0x000E 0x02 0x0004 Used Ignored
0x00
Bytes 8-15
64-bit recording
delay, in scans.
The most
significant 3 bytes
will be ignored
and should be
0x000000.
Specifies the number of scans to delay, after the start command is issued, before beginning time-based
recording. A value of 0 indicates no delay.
Range: 0 to 1099511627775
Default: 0
7.4.2.6 Set Time-Based Recording Skip Count Command
Length
Command
Group
Command
Code
Card Mask Channel
Mask
Parameters
0x000F 0x02 0x0005 Used Ignored
0x00
Byte 8:
Group number
Multiple groups can be
specified by logically ORing
the values:
0x01
Group A
0x02 Group B
0x04 Group C
0x08 Group D
Bytes 9-16:
64-bit skip count, in scans.
The most significant 3 bytes
will be ignored and should be
0x000000.
Determines the number of scans to skip while using time-based recording. A skip factor of 1 will skip
every other point (decimation factor of 2). Specifying a 0 indicates that no scans will be skipped.
Range: 0 to 1099511627775
Default: 0