Vishay Micro-Measurements System 7000 Programmer’s Reference Manual
Page 116 of 142
7.4.3.3 AutoStop Command
Length
Command
Group
Command
Code
Card Mask Channel
Mask
Parameters
0x000E 0x03 0x0003 Used Ignored
Byte 8-15:
64-bit number of scans. The
most significant 3 bytes will
be ignored and should be
0x000000.
Determines the number of scans to acquire before terminating scanning. A value of zero means the
system will not auto stop. The autostop count should be the same for all cards.
7.4.3.4 Get Last Data File Info
Length
Command
Group
Command
Code
Card Mask Channel
Mask
Parameters
0x0006 0x03 0x8004
Query Only
Used Ignored
0x00
Query Returns:
Bytes 8-19:
Last Data File Name in
8.3 format e.g.
XXXXXXXX.7KD
Byte 20:
Null Terminator (0x00)
Bytes 21-24:
File Size in Bytes
Bytes 25-29:
Number of Scans Recorded
(40-bit long integer)
Bytes 30-48
Recording Start Time:
MM/DD/YYYY HH:MM:SS
(note the space between the
year and the hour)
Byte 49:
Null Terminator (0x00)
This function returns the last data file name, size, number of scans recorded, and recording start time.