- 39 -
new calibration coefficients in the internal flash memory, along
with the calibration date.
NOTE: The model P3 is a ratiometric measuring device, so the
measurement is not affected by fluctuations in the excitation
voltage. It is therefore important to use a Wheatstone Bridge-
based calibrator. Other sources, such as a precision voltage ref-
erence, are not suitable for calibration purposes.
4.5.15 Erase MMC
The Erase MMC option erases all files on the multimedia card.
* ADVANCED OPTIONS *
Calibrate System
Upgrade Firmware
Erase MMC
Poisson Ratio: 0.30
Factory Defaults
Exit
Top prevent accidental erasure, a confirmation screen is dis-
played:
Erase MMC?
REC=Yes Menu=No