EX10xxA Function Calls 213
vtex10xxA_self_cal_clear_stored
FUNCTION PROTOTYPE
vtex1048_self_cal_clear_stored(ViSession vi);
FUNCTION PARAMETERS
vi = contains a session handle to the instrument. This handle is obtained by the function and remains valid until the
session is closed.
DATA ITEM RESET VALUE
Not applicable to this function.
DESCRIPTION
This function clears self-calibration data from nonvolatile memory. This operation does not clear the current self-
calibration data, only that in non-volatile memory. If this function is called and no non-volatile self-calibration data
is present, an error is generated.
EXAMPLE