www.vtiinstruments.com
EX10xxA Calibration 303
SECTION 10
EX10XXA/EX10XXA-51C/RX10XX/EX1044
CALIBRATION
INTRODUCTION
It is recommended that the EX10xxA/EX10xxA-51C/RX10xx/EX1044 have factory calibration
performed annually to ensure that the instrument maintains its accuracy and precision. The
following procedure describes the resources and steps necessary to perform factory calibration and
is valid for the following assemblies:
70-0355-100: EX1000A, 48-Ch Voltage Input Instrument
70-0355-200: EX1016A, 32-Ch Voltage, 16-Ch TC Input Instrument
70-0355-300: EX1032A, 16-Ch Voltage, 32-Ch TC Input Instrument
70-0355-400: EX1048A, 48-Ch TC Input Instrument
70-0355-500: EX1000A-TC, 48-Ch Voltage Input Instrument with TC Connectors
70-0564-000: EX1000A-51C, 51-Ch Voltage Input Instrument
70-0536-000: RX1032, 32-Ch Thermocouple/Voltage Instrument
70-0355-451: EX1044, 48-Ch Thermocouple/Voltage/ Current (Isolated) Instrument
REQUIRED RESOURCES
Equipment Needed
LAN/GPIB Gateway (Agilent E5810A recommended)
DMM (either an Agilent 3458A or Keithley 2002)
Test leads with banana jacks
One Ethernet switch
Three Ethernet cables (straight cables)
One GPIB cable
Software Needed
A web browser (Internet Explorer, Firefox, etc.)
The latest version of EX10xxA firmware, available on the VTI Instruments web site.
(The latest version of RX10xx firmware, available on the VTI Instruments web site).
Bonjour (optional: used to discover the EX10xxA/RX10xx, requires use of IE Version 6 or higher)
Agilent Connection Expert (optional: used for instrument discovery if Bonjour is not used)
Any NFS Server software (optional: used to collect the calibration log and data)