EasyManua.ls Logo

VTI Instruments EX1000A - Voltage Connections

VTI Instruments EX1000A
320 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
VTI Instruments Corp.
52 EX10xxA Preparation for Use
Voltage Connections
The EX1000A, EX1016A and EX1032A have 50-pin D-sub receptacles for the voltage input
channels. These channels are differential inputs and are mapped as pairs (see Table 2-3). The center
row of each 50-pin D-sub is connected to chassis GND. This allows for optimized shield termination
when using shielded pair wiring.
P1 Connector
P2 Connector
P3 Connector
Pin
Signal
Pin
Signal
Pin
Signal
1
CJC0_RETURN
1
CJC4_RETURN
1
CJC8_RETURN
2
CH07-HI
2
CH23-HI
2
CH39-HI
3
CH07-LO
3
CH23-LO
3
CH39-LO
4
CH06-HI
4
CH22-HI
4
CH38-HI
5
CH06-LO
5
CH22-LO
5
CH38-LO
6
CH05-HI
6
CH21-HI
6
CH37-HI
7
CH05-LO
7
CH21-LO
7
CH37-LO
8
CH04-HI
8
CH20-HI
8
CH36-HI
9
CH04-LO
9
CH20-LO
9
CH36-LO
10
CH03-HI
10
CH19-HI
10
CH35-HI
11
CH03-LO
11
CH19-LO
11
CH35-LO
12
CH02-HI
12
CH18-HI
12
CH34-HI
13
CH02-LO
13
CH18-LO
13
CH34-LO
14
CH01-HI
14
CH17-HI
14
CH33-HI
15
CH01-LO
15
CH17-LO
15
CH33-LO
16
CH00-HI
16
CH16-HI
16
CH32-HI
17
CH00-LO
17
CH16-LO
17
CH32-LO
34
CJC0_INPUT
34
CJC4_INPUT
34
CJC8_INPUT
35
CH15-LO
35
CH31-LO
35
CH47-LO
36
CH15-HI
36
CH31-HI
36
CH47-HI
37
CH14-LO
37
CH30-LO
37
CH46-LO
38
CH14-HI
38
CH30-HI
38
CH46-HI
39
CH13-LO
39
CH29-LO
39
CH45-LO
40
CH13-HI
40
CH29-HI
40
CH45-HI
41
CH12-LO
41
CH28-LO
41
CH44-LO
42
CH12-HI
42
CH28-HI
42
CH44-HI
43
CH11-LO
43
CH27-LO
43
CH43-LO
44
CH11-HI
44
CH27-HI
44
CH43-HI
45
CH10-LO
45
CH26-LO
45
CH42-LO
46
CH10-HI
46
CH26-HI
46
CH42-HI
47
CH09-LO
47
CH25-LO
47
CH41-LO
48
CH09-HI
48
CH25-HI
48
CH41-HI
49
CH08-LO
49
CH24-LO
49
CH40-LO
50
CH08-HI
50
CH24-HI
50
CH40-HI
TABLE 2-3: VOLTAGE CONNECTOR SIGNAL ASSIGNMENTS
Pin 1
Pin 34
Pin 18
Pin 17
Pin 50
Pin 33
FIGURE 2-12: VOLTAGE CONNECTOR PIN LOCATIONS
Pins 1 and 34 are used for connecting a remote thermistor to the internal signal conditioning circuits.
This is used when an external cold plate is being used for temperature measurement. The
recommended thermistor is U.S. Sensor P/N: PS103J2 or PS103J2-RoHS. This device is 10 kΩ at
25 °C with ±0.1 ºC initial accuracy and the 0 °C to 50 °C BETA is 3890. The thermistor path, within
the EX10xxA product, is single ended and designed to be connected to an electrically isolated
thermistor. The use of a grounded thermistor is not recommended.

Table of Contents

Related product manuals