VTI Instruments Corp.
284 EX10SC Signal Conditioning
NOTES It is important to note that, if connected to a high energy source, damage to the EX10xxA and/or
EX10SC systems and the plug-ins installed can occur. To avoid damage due to high energy sources,
the following precautions should be made:
1) Limit the input current and/or power to the values listed in the plug-ins maximum
specifications.
2) Employ current/power limiting devices (such as fuses) to ensure that excessive current/power
is mitigated in the event a system fault occurs.
CONNECTING TO AN EX10XXA TO THE EX10SC
The EX10SC output signals are designed to be routed directly to the input channels of the EX1000A,
EX1016A or EX1032A via a D-sub cable. The cable (VTI P/N 70-0397-000) directly connects the
EX10SC outputs to the user selected input D-sub connector on either the EX1000A, EX1016A, or
EX1032A.
Connector Pin/Signal Assignment
The table below provide signal and connector pin information for the interconnect cable. For mating
connector information, please refer to the Accessories section of the EX10SC General
Specifications.