EX10xxA Calibration 307
2) Set the Time Source to “Manual”.
3) Enable the Set Time checkbox.
4) Enter the date in the Date (MM:DD:YYYY) fields.
5) Enter the time in time in the Time (HH:MM:SS) fields using a 24-hour format.
6) Click the Submit button.
6. Click submit at last
3. Enable Set Time
2. Set to Manual
1. Select a Time Zone
4. Set the date in MM:DD:YYYY format
5. Set the time in 24-hrs format
FIGURE 10-5: TIME CONFIGURATION
PERFORMING CALIBRATION
Once calibration equipment is connected to the EX10xxA/RX10xx and after ensuring that the time
is set properly, it can be calibrated using the SFP. The following description provides how this is
done.
1) In the browser’s address bar, replace ‘index.cgi?’ with ‘factory.cgi?’ then press the Enter
button. This will bring up the Factory Mode Enable page. Refer to the Login into SFP
discussion for assistance in logging into the EX10xxA/RX10xx web page.
NOTE 1) A warning may appear indicating that a factory calibration file already exists. This warning
can be ignored.
2) Calibration requires a one hour warm-up period. If the EX10xxA/RX10xx not has been on for
one hour, a warning may appear and the calibration process will stop at this point.
2) When the Enable Factory Mode page loads, click the Enable button. The page will indicate
that factory mode is enabled.