268 EX10xxA/RX1032/EX1044 Function Calls
vtex10xxA_set_vtb_wiredor_bias
FUNCTION PROTOTYPE
ViStatus vtex10xxA_set_vtb_wiredor_bias (ViSession vi, ViInt32 wiredor_bias);
FUNCTION PARAMETERS
vi = contains a session handle to the instrument. This handle is obtained by the function and remains valid until the
session is closed.
wiredor_bias = an integer input value that indicates the Wired-OR bias enabled state for all eight LXI Trigger Bus
lines. Only the least significant bits are used. The least significant bit of the integer represents LXI Trigger Bus
line 0, where the 8
th
least significant bit corresponds to LXI Trigger Bus line 7. A value of “1” value for any of the
eight bits indicates that the wired or bias mode is enabled for the particular LXI trigger bus line, while a “0”
indicates that it is disabled.
DATA ITEM RESET VALUE
wiredor_bias = 0
DESCRIPTION
This function sets the Wired-OR bias enable state for each LXI Trigger Bus lines.