304 EX10xxA Calibration
SETTING UP THE INSTRUMENTS
Prior to performing calibration, the EX10xxA/RX10xx must be connected to the required equipment
as indicated in the steps below:
Fig 10-1, Demonstrates the EX10xxA unit for EX10xx calibration
Fig 10-2, Demonstrates the RX10xx unit for RX10xx calibration
1) Connect a GPIB cable from the LAN/GPIB gateway to the DMM.
2) Connect a banana jack probe from the Calibration Sense HI/LO jacks of EX10xxA/RX10xx to
the HI/LO voltage terminals of DMM.
3) Connect an Ethernet cable from the PC’s LAN port to the Ethernet switch.
4) Connect an Ethernet cable from the Ethernet switch to the LAN/GPIB gateway.
5) Power up all the instruments.
6) The GPIB/LAN Gateway should be able to query and list the attached GPIB devices through
the browser. Note the GPIB address of connected DMM.