www.vtiinstruments.com
EX10xxA Triggering 65
channel is linked as a DIO Limit Event, this is noted in the direct control mechanism as a pseudo-
warning to guard against accidental use.
The default selections for each DIO channel are:
output enable is off
output level is 0
The default selections for DIO Limit Events are:
all channels are disabled
invert is off
latch is off
The electrical specifications for the digital I/O port are provided in Table 3-8. Particular note should
be given to the V
INPUT
specification of -0.5 V to 5.5 V. Exceeding that value with an external voltage
source, even though a resistance, could permanently damage the EX10xxA.
TABLE 3-8: DIGITAL I/O PORT ELECTRICAL SPECIFICATIONS
LXI TRIGGER BUS
The EX10xxA series provides an LXI compatible trigger bus connector set on the rear panel of the
instrument. Both connectors have the same pin assignments and are wired in parallel. For more
information on the LXI Trigger Bus and creating an appropriate cable, please visit
www.lxistandard.org and refer to LXI Standard Revision 1.2 and LXI Trigger Bus Cable and
Terminator Specifications Rev 1.1.
1 2 3 4 5 6 7 8 9 10 11 12 13
14 15 16 17 18 19 20 21 22 23 24 25
TABLE 3-9: TRIGGER BUS CONNECTOR PIN ASSIGNMENT