EasyManua.ls Logo

VTI Instruments EX1000A - Page 294

VTI Instruments EX1000A
320 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
VTI Instruments Corp.
294 EX10SC Signal Conditioning
EX10SC-8B34-04 SPECIFICATIONS
DESCRIPTION
2- and 3-wire 100 Ω RTD (0 °C to 600 °C)
INPUT RANGE LIMITS
Input Range
Accuracy
0 °C t o +600 °C (+32 °F to +1112 °F)
±0.45 °C
INPUT RESISTANCE
Normal
Power Off
Overload
50
200 kΩ
200 kΩ
INPUT PROTECTION
Continuous
Transient
240 V ac
ANSI/IEEE C37.90.1
SENSOR EXCITATION CURRENT
0.25 mA
LEAD RESISTANCE EFFECT
±0.02 °C/Ω
CMV (COMMON MODE VOLTAGE)
Input to Output
1500 V rms maximum
TRANSIENT
Input to Output
ANSI/IEEE C37.90.1
CMR (COMMON MODE REJECTION)
50 Hz or 60Hz
120 dB
NMR (NORMAL MODE REJECTION)
70 dB at 60 Hz
ACCURACY
See ordering information
STABILITY
Offset
Gain
±20 ppm/°C
±50 ppm/°C
NOISE
Output, 100 kHz
200 μV rms
BANDWIDTH (-3 dB)
3 Hz
RESPONSE TIME
90% span
150 ms
RTD STANDARDS
100 Ω Pt
Alpha Coefficient
DIN
JIS
IEC
0.00385
DIN 43760
JIS C 1604-1989
IEC 751

Table of Contents

Related product manuals