Preparing the 515 pump 3-19
parameter value is correct.
d. Press Menu to store the parameter.
3. Press Menu until the HPL parameter appears. Press Edit/Enter, then
use the Up or Down key to increase or decrease the value of the
parameter. Note that the increment of change varies according to the
selected Unit parameter.
4. Press Menu until the LPL parameter appears. Press Edit/Enter, then
use the Up or Down key to increase or decrease the value of the
parameter. Note that the increment of change varies according to the
selected Unit parameter.
5. Press Menu until the Mode parameter appears. If required, press the
Edit/Enter key, then the Up or Down key to set the parameter to Local.
Press Menu to store the parameter setting.
6. Press Run/Stop to start the pump.
7. Equilibrate the pump for at least 10 column volumes.
8. When you finish, press Run/Stop to stop the pump.
Equilibrating for remote operation
Caution: Be sure that the entered values for Flow, HPL
(high-pressure limit), and LPL (low-pressure limit) are
appropriate for your column. Some columns require a flow ramp
up. See the Care and Use Manual for your column.
Caution: If your 515 pump is connected to a Waters 680 Automated
Gradient Controller, do not power off the controller if the pump is
running. Doing so may damage the column due to a spurious pressure
output signal from the controller when it is powered off. To avoid this
situation, power off the pump before you power off the controller.