EasyManua.ls Logo

Waters ACQUITY UPLC M-Class - Page 59

Waters ACQUITY UPLC M-Class
117 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Table 4–1: ACQUITY UPLC M-Class system maintenance procedures
Maintenance procedure Frequency
Perform leak test As needed. For instructions on evaluating the
fluid-handling integrity of the system, see the
online Help regarding the following diagnostic
tests:
Fluid integrity test
Dynamic leak test
Static decay test
Set-pressure diagnostic test
Tightening of fittings As needed. For instructions on tightening
fittings, see the component’s overview and
maintenance guide on the documentation
CD.
Replace tubing As needed. For instructions on replacing
tubing, see the component’s overview and
maintenance guide on the documentation
CD.
Table 4–2: Solvent manager maintenance procedures
Maintenance procedure Frequency
Replace the pump interconnect tubing As part of the annual scheduled routine
maintenance and thereafter, as needed. The
typical failure modes are as follows:
You see a visible leak.
The system exhibits a high-pressure
shutdown and the pressure decays when
the connection after the tubing is broken.
November 26, 2019, 715003588 Revision C
Page 59

Table of Contents

Related product manuals