EasyManua.ls Logo

Waters ACQUITY UPLC M-Class - 2.2 General guidelines for a microscale system

Waters ACQUITY UPLC M-Class
117 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Note: If the changes in lab temperature are too large, consider relocating the system to a more
temperature-stable area of the lab, or contact the lab facilities manager to improve lab
temperature stability.
2.2
General guidelines for a microscale system
Separations performed on the ACQUITY UPLC M-Class system use smaller ID columns when
compared to an analytical UPLC system. Solvent and sample consumption are significantly
reduced.
The microscale chromatograph requires optimal performance from the µSM-FL because sample
dispersion is more evident when using smaller columns.
Recommendation: Regarding fast, microscale analyses, a peak of interest can be less than
0.5 seconds wide. For good quantitation, a sampling rate of 10 to 15 points across a peak is
desirable.
2.2.1
Required solvents
Important: Use only high-quality solvents.
See also: Appendix D: Solvent considerations
Requirements:
Use high-quality solvents, buffers, and additives (LC/MS-grade, pre-filtered).
Use high-quality water (LC/MS-grade).
Follow these general recommendations when performing a microscale analysis:
Always use solvent filters on tubing lines in solvent bottles.
Filter salt-containing buffers with a 0.2-µm filter membrane.
Keep concentrated stock solutions to use when you prepare working solutions.
Make fresh buffer solutions daily.
To avoid promoting microbial growth, avoid adding fresh buffer to old buffer.
Keep the seal wash line primed.
Prime solvent lines during system start-up.
Keep all solvent lines primed.
Do not block the degasser vent line. Trim the tubing, if necessary.
Note: Eluent lines A1 and B1 do not pass through the degassers; however the washes are
degassed. When working at flow rates less than 1µL/min, passing the eluent through the
degasser may cause depletion of any volatile mobile phase modifier. For most applications, a
November 26, 2019, 715003588 Revision C
Page 25

Table of Contents

Related product manuals