EasyManua.ls Logo

Waters ACQUITY UPLC M-Class - Troubleshooting M-Class System Devices

Waters ACQUITY UPLC M-Class
117 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Consider the likely causes of the problem. For example, unresponsive modules can mean that
power or signal cables are disconnected or improperly connected. A fluid or vacuum leak can
indicate defective tubing or valve connections.
Look for the less obvious causes of a problem:
Confirm that the system back pressure is appropriate.
Determine the flow rate using a graduated cylinder.
Determine gradient accuracy.
Perform a column efficiency test.
Define problems in detail, for example:
Complete failure of analysis.
Peaks incorrectly identified.
Quantitative problems (bad precision or accuracy).
Run a standard exactly as it was run at installation:
Use a fresh, tested column.
Use bottled, tested eluents.
Adhere to these guidelines when evaluating results of the standard runs:
Retention time reproducibility must be comparable to the results obtained during
installation (the standard deviation maximum is less than 0.25 min for six injections).
Retention times must match those specified for the method.
Peak areas must be comparable to the results obtained during installation.
4.3.2
Troubleshooting M-Class system devices
The following sections describe the most common chromatography, method parameters, and
sample problems defined at a system level.
4.3.2.1
Baseline or background fluctuations
Suspected cause Possible solution
Solvent contamination Use only fresh, clean solvents.
See also: Controlling Contamination in Ultra
Performance LC/MS and HPLC/MS Systems
on the documentation CD.
Erratic solvent manager pressure Prime the µBSM and ASM. If the problem
persists, troubleshoot the solvent manager
pressure.
November 26, 2019, 715003588 Revision C
Page 64

Table of Contents

Related product manuals