Suspected cause Possible solution
Loose fittings or crimped lines before or after
the column
Inspect all fittings and tubing. Replace as
necessary.
Inadequate solvent blending If the baseline is steady when running a
premixed solvent, then:
• Ensure that A and B pumps are delivering
solvent correctly.
• For shallow gradients or protein
applications running at lower pressure,
switch to the mixer kit for the 1-mm column.
NanoFlow sprayer/emitter problem Clean or replace the emitter.
See also: Universal NanoFlow Sprayer
Installation and Maintenance Guide on the
documentation CD.
Increased background caused by possible:
• Dirty source
• Detector setting too high
• Electronic noise
• Column problem
• Solvent clusters
Ensure cleanliness of solvents and proper use
of the column:
• Adjust the detector setting and refer to the
appropriate documentation for that
instrument.
• Clean the source and sonicate the sample
cone in 50:50 methanol:water.
• Optimize the source conditions.
Dirty flow cell Ultraviolet (UV), system only Clean the flow cell.