Symbol Definition
Alternating current
Electrical and electronic equipment with this
symbol may contain hazardous substances
and should not be disposed of as general
waste
For compliance with the Waste Electrical and
Electronic Equipment Directive (WEEE)
2012/19/EU, contact Waters Corporation for
the correct disposal and recycling instructions
For indoor use only
No pushing
10kg
max
Indicates the maximum load you can place on
that item (for example, 10kg)
Serial number
REF
Part number, catalog number
Audience and purpose
This guide is intended for personnel who install, operate, and maintain ACQUITY UPLC M-Class
system modules. It gives an overview of the system’s technology and operation.
Intended use of the ACQUITY UPLC M-Class system
Waters designed the ACQUITY UPLC M-Class system to perform chromatographic separations.
The system is for research use only and is not intended for use in diagnostic applications.
Calibrating
To calibrate LC systems, adopt acceptable calibration methods using at least five standards to
generate a standard curve. The concentration range for standards must include the entire range
of QC samples, typical specimens, and atypical specimens.
November 26, 2019, 715003588 Revision C
Page vi