EasyManua.ls Logo

Waters ACQUITY UPLC M-Class - Page 66

Waters ACQUITY UPLC M-Class
117 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4.3.2.2 Erratic retention times
Suspected cause Possible solution
The system is not stabilized or chemically
equilibrated
Allow all system components (such as the
column and detector) sufficient time to
stabilize and chemically equilibrate.
Ensure proper operating conditions for the
application (such as mobile phase, detector
settings, and detector type).
Verify the initial hold at the beginning of the
gradient.
Ensure that the flow controllers were
autozeroed with the solvent currently in use
at a temperature close to the current lab
temperature.
See also: The component’s overview and
maintenance guide, or the column care and
use manual, on the documentation CD for
recommended equilibration times.
Note: When running an automated gradient
method, ensure that there is sufficient and
reproducible equilibration times used between
injections.
Erratic solvent manager pressure Prime the solvents.
Run leak tests.
Analyze the pressure trace in the Console
for stability.
Injection volume or sample concentration is too
high; disrupting equilibrium
Reduce the injection volume or dilute the
sample with initial mobile phase.
Ambient temperature fluctuations Stabilize the operating environment
temperature (for a minimum of 30 minutes).
If the problem persists:
Run the column heater temperature at 5 °C
above ambient.
Close any open air vents.
Contact your Waters service
representative.
November 26, 2019, 715003588 Revision C
Page 66

Table of Contents

Related product manuals