Suspected cause Possible solution
Inconsistent retention times • Flush the column with 85% organic solvent.
• Clean or replace the column as described
in the appropriate column care and use
manual on the documentation CD.
• Rerun the analysis and observe whether
the retention times stabilize.
• Ensure cleanliness of the mobile phase.
4.3.2.3
Increasing or decreasing retention times
Suspected cause Possible solution
Decreased flow rate Verify the flow rate setting.
See also: The Universal NanoFlow Sprayer
Installation and Maintenance Guide on the
documentation CD.
Incorrect solvent composition Change the solvent composition, and then
autozero.
Incorrect, degraded, or contaminated column • If contaminated, flush the column with 85%
organic solvent.
• Otherwise, clean or replace the column as
described in the appropriate columns care
and use manual on the documentation CD.
The system is not stabilized or chemically
equilibrated
• Allow all system components (such as the
column and detector) sufficient time to
stabilize and chemically equilibrate.
• Ensure proper operating conditions for the
application (such as mobile phase, detector
settings, and detector type).
See also: The component’s overview and
maintenance guide or the column care and use
manual on the documentation CD for
recommended equilibration times.
Note: When running an automated gradient
method, ensure that there is sufficient and
reproducible equilibration times between
injections.