Starting the TQ Detector 2-3
4. Press the power switch on the top, left-hand side of the TQ Detector and
ACQUITY instruments. Each system instrument “beeps” and runs a
series of startup tests.
5. Allow 3 minutes for the embedded PC to initialize. An audible alert
sounds when the PC is ready.
The power and status LEDs change as follows:
Each system instrument’s power LED shows green.
During initialization, the binary solvent manager’s and sample
manager’s status LED flashes green.
After the instruments are successfully powered-on, all power LEDs
show steady green. The binary solvent manager’s flow LED, the
sample manager’s run LED, and the TQ Detector’s operate LED
remain off.
6. Start MassLynx. You can monitor the ACQUITY console for messages
and LED indications.
7. In the MassLynx main window’s lower-left corner, click IntelliStart.
Result: TQ Detector console IntelliStart appears. The TQ Detector is in
Standby mode.