Installing the corona pin 3-5
Installing the corona pin
Required materials
Chemical-resistant, powder-free gloves
Needle-nose pliers
80:20 acetonitrile/water
To install the corona pin:
1. In the ACQUITY UPLC Console, click Standby , and confirm that
the Operate indicator is not illuminated.
2. Open the instrument’s access door.
3. Unlatch the source enclosure door’s handle and open the door.
Warning: The ACQUITY UPLC system connections, ESI probe,
and source can be contaminated with biohazardous and/or toxic
materials. Always wear chemical-resistant, powder-free gloves
while performing this procedure.
Warning: To avoid electric shock, ensure that the instrument is in
Standby mode when commencing this procedure.
Warning: The source can be hot. To avoid burn injuries, take
great care while working with the instrument’s access door open.
Warning: The probe tip is sharp. To avoid puncture wounds, take
great care while working with the source enclosure door open if
an ESI probe is fitted.
Caution: Do not apply any downward force to the source
enclosure door while the door is open.

Table of Contents

Other manuals for Waters TQ Detector

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Waters TQ Detector and is the answer not in the manual?

Waters TQ Detector Specifications

General IconGeneral
BrandWaters
ModelTQ Detector
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals