v
Flammable solvents hazard
Never let the nitrogen supply pressure fall below 400 kPa (4 bar, 58 psi)
during analyses that require flammable solvents. Connect to the LC output
with a gas-fail connector to stop the LC solvent if the nitrogen supply fails.
High temperature hazard
Waters TQ Detector high temperature hazard:
Warning: Where significant quantities of flammable solvents are
involved, the source must receive a continuous flow of nitrogen to
prevent possible ignition within that enclosed space.
Warning: The source enclosure can be hot. To avoid burn injuries,
avoid touching the source enclosure when operating or servicing the
TQ Detector.
Source enclosure assembly