Performing a sample tune 4-7
Example ACQUITY UPLC Console display during calibration:
IntelliStart creates tune and calibration files, which are saved as
specified on the Instrument Setup Parameters dialog box’s Instrument
Setup tab. Once calibration is complete, the sample tune starts on the
four masses defined in the IntelliStart Setup Parameters dialog box.
When the sample tune is complete, the results are written to the
Sulfadimethoxine.ipr file.
Performing a sample tune
In the following example, sulfadimethoxine is used as the tuning sample.
Required materials
Sulfadimethoxine solution (100 pg/µL)
To perform a sample tune:
1. Prepare the IntelliStart fluidics system with sulfadimethoxine solution
in reservoir B.
See also: “To prepare the IntelliStart fluidics system:” on page 4-2.