EasyManua.ls Logo

Waters TQ Detector - Page 72

Default Icon
92 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4-8 Operating the Waters TQ Detector
2. In the ACQUITY UPLC Console system tree, click TQ Detector.
3. Ensure that the Ion Mode is ES+.
4. In the ACQUITY UPLC Console system tree, click IntelliStart.
5. Clear the Instrument Setup check box.
6. If you require system pre-checking, click Pre-checks.
Rationale: If Pre-checks is selected, when IntelliStart starts the sample
tune, it determines whether the existing tune is still valid. If so, no
sample tune is performed.
See also: TQ Detector online Help topic “IntelliStart flow diagram”.
7. Click Start .
8. In the IntelliStart Setup Parameters dialog box’s Sample Tune tab,
specify the sample tune parameters as described in “To specify the
sample tune parameters:” on page 4-5.
9. Click Start .
Result: A message appears reminding you to ensure that the tune
solutions, tune parameters, and LC flow are set correctly.
10. Click OK.
Result: Sample tune starts. The ACQUITY UPLC Console displays the
progress of the tune. When the sample tune is complete, the results are
written to the Sulfadimethoxine.ipr file.

Table of Contents

Other manuals for Waters TQ Detector

Related product manuals