3-10 ESI and ESCi Modes of Operation
4. Use the needle-nose pliers to remove the corona pin from its mounting
contact.
Tip: Store the corona pin in a safe location.
See also: Figure titled “Corona pin:” on page 3-7.
5. Use the needle-nose pliers to fit the blanking plug to the corona pin
mounting contact.
See also: Figure titled “Corona pin mounting contact:” on page 3-6.
6. Close the source enclosure door and fasten the handle.
7. Close the instrument’s access door.
Warning: The corona pin tip is sharp. To avoid puncture wounds,
handle the corona pin with care.
Caution: To avoid damaging to the corona pin’s tip and bending
the pin, use the needle nose pliers to grip the corona pin at the
end that fits into the mounting contact.