EasyManua.ls Logo

Xo FLEX - Implant Surgery - Operation with XO Smart Link; Air Instrument

Xo FLEX
95 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
XO FLEX Instructions for use Operation
25
The motor operates at maximum torque.
Please see section 3.11.3 if you need to control the maximum torque!
IMPLANT SURGERY - OPERATION WITH XO SMART LINK
In combination with XO Smart Link you can work with instrument programs and implant
treatments.
Please see “XO Smart Link Instructions for use”.
3.12 AIR INSTRUMENT
An air turbine or an air scaler may be attached to the hose in the following called “the air
instrument”.
The air instrument is used as described by the supplier.
Air instruments with a type 3 coupling as specified in ISO 9168 shall be used.
Lift the air instrument forward and control it with the foot control.
Activate the air instrument (one step) with or .
Figure 23Foot control of turbine
See drive air (in % of maximum) on the display.
Change spray selection (water & air, air only or no spray) with X.
Enable/disable automatic chip blow with O.
When the automatic chip blow is enabled a short burst of high-pressure air dries the
preparation each time an instrument with spray water stops.
Using the automatic chip blow function significantly reduces the number of shifts
between rotation instrument and syringe.
X
O

Table of Contents

Other manuals for Xo FLEX

Related product manuals