7 UNIT MESSAGES AND REMEDIAL ACTIONS
7.1 ERROR MESSAGES
Certain functional errors can be detected by XO FLEX and written onto the instrument bridge
display.
Table 8 – Error messages
CALL SERVICE
Displayed after start-up of unit if
no communication with foot
control is detected.
Call your XO authorized service
provider.
CALL SERVICE
Displayed after start-up of unit if
no communication with Stand
Controller and foot control is
detected.
Call your XO authorized service
provider.
Displayed when unit detects that
it takes more than 90 seconds to
fill the mixing cup.
Check the water supply to the
unit or call your XO authorized
service provider.
CALL SERVICE
Displayed after start-up of unit if
no communication with Power
Driver is detected.
Call your XO authorized service
provider.
CALL SERVICE
Displayed after start-up of unit if
no communication with Stand
Controller is detected.
Call your XO authorized service
provider.
CALL SERVICE
Displayed when unit detects that
it takes more than 90 seconds to
fill the mixing cup or mixing cup
cannot be emptied within 4
minutes.
Call your XO authorized service
provider.
Displayed if unit detects that the
foot control is activated – or an
instrument is lifted forward during
start-up.
Switch off the unit and try again.
If this does not help, call your XO
authorized service provider.
Patient chair lifting motor is
overheated. Do not adjust the
height or inclination of the chair
for 10 minutes and try again.
If this does not help, contact your
XO authorized service provider.
CALL SERVICE
Water leakage. Switch off water
supply and try to localize
leakage.
If necessary, contact your XO
authorized service provider.
One or more unit instruments are
not fitted correctly. Check that
numbers on instrument hoses fit
correct positions at instrument
bridge – see
Figure 18, page 21.
Another possible error is that a
unit instrument hose is defective.
If necessary, contact your XO
authorized service provider.