XO FLEX Instructions for use Installation
8
2 INSTALLATION
XO FLEX is intended to be permanently installed in a dental operatory at least 220 cm wide and
360 cm long
1
see Figure 1 and see the required operation conditions in Table 1.
For transport and storage conditions see Table 1 and in Table 2 the installation requirements are
given.
Figure 1 – Plan of installation XO FLEX
Table 1 – Operating and transport/storage conditions
Condition
Operation
Transport and storage
Temperature:
+15
o
C – +40
o
C
-40
o
C – +60
o
C
Relative humidity:
20% – 85%
10% – 95%
Air pressure:
800 hPa – 1060 hPa
700 hPa – 1060 hPa
Installation altitude
Max. 2,000 meters above sea level
-
1
Note: This dimension depends on the depth of the cabinets in the example shown in Figure 1 the cabinets are 50 cm deep