XO FLEX Instructions for use Cleaning and infection control
58
Figure 72Removing the light rod from XO ODONTOCURE handpiece
5.14 INFECTION CONTROL OTHER INSTRUMENTS
For turbines, other ultrasonic scalers, intraoral video cameras and other instruments please refer to
the user manual supplied with the instrument.
5.15 INFECTION CONTROL XO PERISTALTIC PUMP
Clean and disinfect the surfaces of the pump unit using a lint-free piece of soft cloth with Dürr FD
333.
Tubing clips and spear can be autoclaved.
Tube guide can be thermo-disinfected.
Risk of contamination!
The pre-sterilized disposable irrigation kits are for single-use only and must be
discarded after use.
The tubing must not be re-sterilized.
5.16 DISINFECTION OF LIGHT
Be sure to switch off the unit before cleaning the light.
The protection screen may be detached for cleaning by removing the two screws see Figure 73.
Figure 73Removing the protection screen, (1) screws and (2) protection screen