XO FLEX Instructions for use Maintenance and repairs
71
To avoid unintended noise from the suction hoses, use each suction hose to empty a
glass of water at the beginning of the day and if necessary during the workday.
Suction hoses and filters should be exchanged every 12 months. Replacement suction filters are
included in the infection control kitssee section 10.3.
Amalgam waste is considered hazardous to the environment and should therefore be
disposed of safely and in accordance with regulatory requirements. Remember to
use safety gloves.
6.12 AIR AND AMALGAM SEPARATOR
DÜRR CS1 COMBI-SEPAMATIC AIR SEPARATOR
Service and replacement of this air separator shall be done by an XO authorized service provider.
DÜRR CAS 1 COMBI-SEPARATOR – AIR AND AMALGAM SEPARATOR
If your XO FLEX unit is equipped with an amalgam separator you need to replace the amalgam
collector vessel every 69 months.
Contact your XO authorized service provider.
A warning signal will be heard from the amalgam separator when the amalgam
collector vessel is 95% full or more.