XO FLEX – Instructions for use Operation
31
Figure 29 – Foot control of XO ODONTOCURE
See elapsed exposure time on the display during the process.
Hear a beep at intervals of 5 seconds.
Never look directly into the light or direct it at the eyes of others!
XO ODONTOCURE emits thermal radiation, and blue light and ultraviolet light in the
range 385 – 515 nm at an intensity that requires protection of the eyes. Protect eyes
with light shield and/or protective eyewear that removes light in the previously
mentioned wavelengths.
CAUTION! The high light intensity of XO ODONTOCURE is accompanied by heat
generation on the exposed surface! Uninterrupted exposure of more than 20 seconds
to the same surface shall be avoided. Curing at intermitted intervals is
recommended.
Switch soft-start on and off with O while the light is not activated.
Soft-start is a feature that can help reduce shrinkage. When enabled, the light will emit light at a
reduced intensity for a manually preset number of seconds before going to full intensity.
For curing of posterior teeth use a protection cap:
Figure 30 Protection cap
For curing of anterior teeth use the light shield: