XO FLEX Instructions for use Maintenance and repairs
68
6. The measured depth of the polymerized material shall be recorded and is now the target
reference for future measurements.
Figure 84Testing device
Figure 85Non-polymerized material is removed from composite test plug
Figure 86Measuring the depth of the test plug
At an interval of approximately 1 month, perform the following steps:
7. Please refer to points 1-5 on page 67.
8. Compare the result of this test with the reference made upon receiving the instrument. If the
curing depth deviates more than 0.8mm from the reference, a fault may be present, and
you might need to contact your XO service provider to remedy the fault.
Note: This is a technical verification of performance; it does not reflect actual curing
depth in a human tooth.