Source
1
2
3
4
5
6
7
8
9
10
11
12
13
App
4.2 Sourcing DC Voltage, DC Current or SINK Current Signal
4.2 Sourcing DC Voltage, DC Current or
SINK Current Signal
4.2.1 Sourcing DC Voltage or DC Current Signal
Step 1: Using the Function selector switch, select the desired source
function from
, , ,
and
.
Step 2: The LCD shows the default value and unit of the source
function.
Step 3: Set the output value digit by digit using each pair of
and
output setting keys.
Each pair of
and keys corresponds to each digit of the
LCD reading. Each press of the
and key increases or
decreases the digit. Increasing the digit from 9 or decreasing
it from 0 causes the digit to overow or underow, allowing
you to set the output value without interruption. Holding down
the
or key continuously changes the digit in question.
Pressing the
key initializes the output setpoint to the
default value (0).
Step 4: Pressing the
key causes the indicator on the LCD
to change from to . The calibrator sources the preset
DC voltage or current signal between the output terminals.
Step 5: To turn off the output, press the
key once again.
The appears on the LCD and the output terminals are
open-circuited.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA CA71 HANDY CAL and is the answer not in the manual?

YOKOGAWA CA71 HANDY CAL Specifications

General IconGeneral
BrandYOKOGAWA
ModelCA71 HANDY CAL
CategoryTest Equipment
LanguageEnglish

Related product manuals