BL
Turns
the
back
lighting
on
and
off
/queries
the
current setting.
When normal
condition
Syntax for setting
BLm<delimiter>
Syntax for query
BL?<delimiter> ⇒ Response: BLm<delimiter>
Description of parameter
m=0: Off m=1: On
DW
Moves down the “m-th” digit of the sourced setpoint
by one digit.
When normal
condition
Syntax for setting
DWm<delimiter>
Description of parameter
m: Specifies a digit 1 (the lowest digit) to 5 (the highest digit)
UP
Moves up the “m - th” digit of the sourced setpoint
by one digit.
When normal
condition
Syntax for setting
UPm<delimiter>
Description of parameter
m: Specifies a digit 1 (the lowest digit) to 5 (the highest digit)
H
Enables/Disables
the
output
data
header/queries
the
current setting.
When normal
condition/adjustment
Syntax for setting
Hm<delimiter>
Syntax for query
H?<delimiter> ⇒ Response: Hm<delimiter>
* For details on the header, see the OD command.
Description of parameter
m: Enables/Disables the header
m=0: Disabled m=1: Enabled
HD
Enables/Disables
data
hold
mode/queries
the
current
setting.
When normal
condition
Syntax for setting
HDm<delimiter>
Syntax for query
HD? <delimiter> ⇒ Response: HDm (delimiter)
Description of parameter
m: Specifies data hold
m=0: Hold Off m=1: Hold On