MF Queries the measurement function.
When normal
condition
Syntax for query
MF? <delimiter> ⇒ Response: MFm<delimiter>
Description of parameter
m: Measurement function
m=0: 300V AC m=1: 100V m=2: 10V
m=3: 1V m=4: 100mV m=5: Resistance
m=6: Frequency m=7: Current
MO On/Off of measurement/queries the current setting.
When normal
condition/adjustment
Syntax for setting
MOm<delimiter>
Syntax for query
MO?<delimiter> ⇒ Response: MOm<delimiter>
Description of parameter
m: On/Off condition
m=0: Off m=1: On
MR
Sets
the
measuring
range/queries
the
current
setting.
Syntax for setting
MRm<delimiter>
Syntax for query
MR? <delimiter> ⇒ Response: MRm<delimiter>
Description of parameter
m: Measuring range
[100V] m=0: DC m=1: AC
[10V] m=0: DC m=1: AC
[1V] m=0: DC m=1: AC
[100mV] (When normal condition)
m=0: 100mV DC m=1: TcK m=2: TcE
m=3: TcJ m=4: TcT m=5: TcR
m=6: TcB m=7: TcS m=8: TcN
m=9: TcL m=10: TcU
[100mV] (When adjustment)
m=0: 100mV DC m=1: TcK m=2: TcE
m=4: TcT
[Ω] m=0: 400Ω m=1: Pt100 m=2: JPt
[Freq ] m=0: 100Hz m=1: 1kHz m=2: 10kHz
[mA]
m=0: 20mA m=1: 100mA
When normal
condition/adjustment