4-10
IM CA71-E
Step 5: Set the output value digit by digit using each pair of
and
output setting keys.
Each pair of
and keys corresponds to each digit of the
LCD reading. Each press of the
or key increases or
decreases the digit. Increasing the digit from 9 or decreasing
it from 0 causes the digit to overow or underow, allowing
you to set the output value without interruption.
Holding down the
or key continuously changes the digit
in question.
Pressing the
key initializes the output setpoint to the
default value (600°C for a type-B thermocouple).
Step 6: Pressing the
key causes the indicator on the LCD
to change from to . A thermoelectromotive force based
on the temperature detected by the RJ sensor develops
between the output terminals.
Step 7: To turn off the output, press the
key once again.
The appears on the LCD and the output terminals are
open-circuited.
NOTE
• When you have attached the RJ sensor to the device being calibrated,
wait until the detected temperature stabilizes before you begin using
the calibrator.
• If no reference junction compensation is required, be sure to remove
the RJ sensor from the calibrator.
4.4 Sourcing Thermocouple (TC) Signals

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA CA71 HANDY CAL and is the answer not in the manual?

YOKOGAWA CA71 HANDY CAL Specifications

General IconGeneral
BrandYOKOGAWA
ModelCA71 HANDY CAL
CategoryTest Equipment
LanguageEnglish

Related product manuals