4-14
IM CA71-E
Step 5: Set the output value digit by digit using each pair of
and
output setting keys.
Each pair of
and keys corresponds to each digit of the
LCD reading. Each press of the
or key increases or
decreases the digit. Increasing the digit from 9 or decreasing
it from 0 causes the digit to overow or underow, allowing
you to set the output value without interruption. Holding down
the
or key continuously changes the digit in question.
Pressing the
key initializes the output setpoint to the
default value (0.1 V).
Step 6: Press the
key once again to show on the LCD.
Then, press the
key one more time to revert to
frequency setting mode.
Step 7: Pressing the
key causes the indicator on the LCD
to change from to . The calibrator sources a continuous
pulse train with the preset frequency and amplitude between
the output terminals.
Step 8: To turn off the output, press the
key once again.
The symbol appears on the LCD and the output terminals
are open-circuited.
TIP
To change the frequency range, place the calibrator in frequency setting
mode with the
key.
Then, change the frequency range using the
key.
4.5 Sourcing Pulse Signals