IM 04L05A01-02E
Confirming the Data
Make sure to check the measured data and alarm information before applying
electronic signature. For the operating procedure of the Viewer, see chapter 3 in
IM04L05A01-61E.
Apply Electronic Signature
1. Click here.
The [Apply signature] dialog box opens.
The next dialog box opens.
The next dialog box opens.
2. Enter the user name, user ID, and password.
3. Click here.
7. Click here.
4. Select the signature No.
5. Select Pass or Fail.
6. Enter comment.
8. Click here.
Note
You cannot apply electronic signature in the following cases.
Signature 1 to 3 are already applied.
Files that make up the batch data is missing.
Continuous data is being displayed by file linking.
File is corrupt.
If you repeat the operation of entering a wrong user ID or password and clicking [OK]
three consecutive times, the user name can no longer be used to apply electronic
signature or display the login information of the data. “Password invalid” is logged in the
operation log list.
If the number of samples in the batch data exceeds 5,000,000, the data is displayed in
blocks of 5,000,000 samples or less on the Viewer. You must display and confirm all
blocks before you can apply electronic signature to the batch data.
Applying Electronic Signature Using DAQSIGNIN

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA DAQSTATION DX100P and is the answer not in the manual?

YOKOGAWA DAQSTATION DX100P Specifications

General IconGeneral
BrandYOKOGAWA
ModelDAQSTATION DX100P
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals