5-97
IM 701310-17E
Commands
1
2
3
4
5
6
App
Index
:GONogo:ZPARameter:SELect<x>:
PARameter:MEASure:TRACe<x>:AREA<x>:
TYPE?
Function Queries all settings related to the waveform
parameters of the measure determination.
Syntax
:GONogo:ZPARameter:SELect<x>:PARameter:
MEASure:TRACe<x>:AREA<x>:TYPE?
<x> of SELect<x> = 1 to 4
<x> of
TRACe<x> = 1 to 8
<x> of AREA<x> = 1 or 2
Example
:GONOGO:ZPARAMETER:SELECT1:PARAMETER:
MEASURE:TRACE1:AREA1:TYPE? -> :GONOGO:
ZPARAMETER:SELECT1:PARAMETER:
MEASURE:TRACE1:AREA1:TYPE:
MAXIMUM 2.000E+00,1.000E+00
:GONogo:ZPARameter:SELect<x>:
PARameter:MEASure:TRACe<x>:AREA<x>:
TYPE:<Parameter>
Function Sets the upper and lower limits of the waveform
parameter of the measure determination or queries
the current setting.
Syntax
:GONogo:ZPARameter:SELect<x>:PARameter:
MEASure:TRACe<x>:AREA<x>:TYPE:
<Parameter> {<NRf>,<NRf>|
<Voltage>,<Voltage>|
<Current>,<Current>|<Time>,<Time>|
<Frequency>,<Frequency>}
:GONogo:ZPARameter:SELect<x>:PARameter:
MEASure:TRACe<x>:AREA<x>:TYPE:
<Parameter>?
<x> of SELect<x> = 1 to 4
<x> of
TRACe<x> = 1 to 8
<x> of AREA<x> = 1 or 2
<Parameter> = {BURSt|CMEan|COUNt|CRMS|
CSDeviation|DELay|DT|DUTYcycle|FALL|
FREQuency|HIGH|HILow|LOW|MAXimum|MEAN|
MINimum|NOVershoot|NWIDth|PERFrequency|
PERiod|POVershoot|PTOPeak|PWIDth|RISE|
RMS|SDEViation|TYCInteg|TYINteg|V1|V2}
<NRf>, <V
oltage>, <Current>, <T
ime>, and
<Frequency> = See the DL9000 User’s Manual.
Example (The following is an example for the maximum value
of trace 1 and area 1.)
:GONOGO:ZPARAMETER:SELECT1:PARAMETER:
MEASURE:TRACE1:AREA1:TYPE:MAXIMUM 1,2
:GONOGO:ZPARAMETER:SELECT1:PARAMETER:
MEASURE:TRACE1:AREA1:TYPE:MAXIMUM?
-> :GONOGO:ZPARAMETER:SELECT1:
PARAMETER:MEASURE:TRACE1:AREA1:TYPE:
MAXIMUM 2.000E+00,1.000E+00
Description
This command applies to cycle statistical processing
when
:MEASURE:MODE CYCLE
.
:GONogo:ZPARameter:SELect<x>:
PARameter:MEASure:STATistics
Function Sets the statistical value of the measure determination
or queries the current setting.
Syntax
:GONogo:ZPARameter:SELect<x>:PARameter:
MEASure:STATistics {MAXimum|MEAN|
MINimum|SIGMa}
:GONogo:ZPARameter:SELect<x>:PARameter:
MEASure:STATistics?
<x> = 1 to 4
Example
:GONOGO:ZPARAMETER:SELECT1:PARAMETER:
MEASURE:STATISTICS MAXIMUM
:GONOGO:ZPARAMETER:SELECT1:PARAMETER:
MEASURE:STATISTICS? ->
:GONOGO:ZPARAMETER:SELECT1:PARAMETER:
MEASURE:STATISTICS MAXIMUM
Description This command is valid when
:MEASURE:MODE
CYCLE
.
:GONogo:ZPARameter:SELect<x>:
PARameter:MEASure:TRACe<x>?
Function Queries all settings related to the trace of the
measure determination.
Syntax
:GONogo:ZPARameter:SELect<x>:PARameter:
MEASure:TRACe<x>?
<x> of SELect<x> = 1 to 4
<x> of
TRACe<x> = 1 to 8
Example
:GONOGO:ZPARAMETER:SELECT1:PARAMETER:
MEASURE:TRACE1? -> :GONOGO:ZPARAMETER:
SELECT1:PARAMETER:
MEASURE:TRACE1:AREA1:TYPE:
MAXIMUM 2.000E+00,1.000E+00
:GONogo:ZPARameter:SELect<x>:
PARameter:MEASure:TRACe<x>:AREA<x>?
Function Queries all settings related to the area of the measure
determination.
Syntax
:GONogo:ZPARameter:SELect<x>:PARameter:
MEASure:TRACe<x>:AREA<x>?
<x> of SELect<x> = 1 to 4
<x> of
TRACe<x> = 1 to 8
<x> of AREA<x> = 1 or 2
Example
:GONOGO:ZPARAMETER:SELECT1:PARAMETER:
MEASURE:TRACE1:AREA1? -> :GONOGO:
ZPARAMETER:SELECT1:PARAMETER:
MEASURE:TRACE1:AREA1:TYPE:
MAXIMUM 2.000E+00,1.000E+00
5.12 GONogo Group

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA DL9040 and is the answer not in the manual?

YOKOGAWA DL9040 Specifications

General IconGeneral
BrandYOKOGAWA
ModelDL9040
CategoryTest Equipment
LanguageEnglish

Related product manuals