5-96
IM 701310-17E
:GONogo:ZPARameter:SELect<x>:
PARameter:FFT<x>:PEAK:DV
Function Sets the upper and lower limits between the peak
voltages of the FFT determination or queries the
current setting.
Syntax
:GONogo:ZPARameter:SELect<x>:PARameter:
FFT<x>:PEAK:DV {<NRf>,<NRf>}
:GONogo:ZPARameter:SELect<x>:PARameter:
FFT<x>:PEAK:DV?
<x> of SELect<x> = 1 to 4
<x> of FFT<x> = 1 or 2
<NRf> = –4 to 4 (div)
Example
:GONOGO:ZPARAMETER:SELECT1:PARAMETER:
FFT1:PEAK:DV 0,1
:GONOGO:ZPARAMETER:SELECT1:PARAMETER:
FFT1:PEAK:DV? -> :GONOGO:ZPARAMETER:
SELECT1:PARAMETER:FFT1:PEAK:
DV 1.000E+00,0.000E+00
:GONogo:ZPARameter:SELect<x>:
PARameter:FFT<x>:PEAK:FREQuency<x>
Function Sets the upper and lower limits of the peak frequency
of the FFT determination or queries the current
setting.
Syntax
:GONogo:ZPARameter:SELect<x>:PARameter:
FFT<x>:PEAK:FREQuency<x> {<Frequency>,
<Frequency>}
:GONogo:ZPARameter:SELect<x>:PARameter:
FFT<x>:PEAK:FREQuency<x>?
<x> of SELect<x> = 1 to 4
<x> of FFT<x> = 1 or 2
<x> of FREQuency<x> = 1 or 2
<Frequency> = See the DL9000 User’s Manual
Example
:GONOGO:ZPARAMETER:SELECT1:PARAMETER:
FFT1:PEAK:FREQUENCY1 0,1
:GONOGO:ZPARAMETER:SELECT1:PARAMETER:
FFT1:PEAK:FREQUENCY1? -> :GONOGO:
ZPARAMETER:SELECT1:PARAMETER:FFT1:
PEAK:FREQUENCY1 1.000E+00,0.000E+00
:GONogo:ZPARameter:SELect<x>:
PARameter:FFT<x>:PEAK:V<x>
Function Sets the upper and lower limits of the peak voltage of
the FFT determination or queries the current setting.
Syntax
:GONogo:ZPARameter:SELect<x>:PARameter:
FFT<x>:PEAK:V<x> {<NRf>,<NRf>}
:GONogo:ZPARameter:SELect<x>:PARameter:
FFT<x>:PEAK:V<x>?
<x> of SELect<x> = 1 to 4
<x> of FFT<x> = 1 or 2
<x> of V<x> = 1 or 2
<NRf> = –4 to 4 (div)
Example
:GONOGO:ZPARAMETER:SELECT1:PARAMETER:
FFT1:PEAK:V1 1,2
:GONOGO:ZPARAMETER:SELECT1:PARAMETER:
FFT1:PEAK:V1? -> :GONOGO:ZPARAMETER:
SELECT1:PARAMETER:FFT1:PEAK:
V1 2.000E+00,1.000E+00
:GONogo:ZPARameter:SELect<x>:
PARameter:MEASure?
Function Queries all settings related to the determination using
automated measurement of waveform parameters
(measure determination).
Syntax
:GONogo:ZPARameter:SELect<x>:PARameter:
MEASure?
<x> = 1 to 4
Example
:GONOGO:ZPARAMETER:SELECT1:PARAMETER:
MEASURE? -> :GONOGO:ZPARAMETER:
SELECT1:PARAMETER:MEASURE:
CALCULATION1 2.000E+00,1.000E+00
:GONogo:ZPARameter:SELect<x>:
PARameter:MEASure:CALCulation<x>
Function Sets the upper and lower limits of the calculation item
of the measure determination or queries the current
setting.
Syntax
:GONogo:ZPARameter:SELect<x>:PARameter:
MEASure:CALCulation<x> {<NRf>,<NRf>}
:GONogo:ZPARameter:SELect<x>:PARameter:
MEASure:CALCulation<x>?
<x> of SELect<x> = 1 to 4
<x> of CALCulation<x> = 1 to 4
<NRf> = –4 to 4 (div)
Example
:GONOGO:ZPARAMETER:SELECT1:PARAMETER:
MEASURE:CALCULATION1 1,2
:GONOGO:ZPARAMETER:SELECT1:PARAMETER:
MEASURE:CALCULATION1? -> :GONOGO:
ZPARAMETER:SELECT1:PARAMETER:MEASURE:
CALCULATION1 2.000E+00,1.000E+00
5.12 GONogo Group

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA DL9040 and is the answer not in the manual?

YOKOGAWA DL9040 Specifications

General IconGeneral
BrandYOKOGAWA
ModelDL9040
CategoryTest Equipment
LanguageEnglish

Related product manuals