5-120
IM 701310-17E
:MATH<x>:POSition
Function Sets the vertical position of the computed waveform
or queries the current setting.
Syntax
:MATH<x>:POSition {<NRf>}
:MATH<x>:POSition?
<x> = 1 to 8
<NRf> = –4 to 4 (div)
Example
:MATH1:POSITION 0
:MATH1:POSITION? -> :MATH1:
POSITION 0.000E+00
:MATH<x>:SCALe?
Function Queries all settings related to scaling.
Syntax
:MATH<x>:SCALe?
<x> = 1 to 8
Example
:MATH1:SCALE? -> :MATH1:SCALE:
CENTER 1.000E+00;MODE AUTO;
SENSITIVITY 1.000E+00
:MATH<x>:SCALe:CENTer
Function Sets the offset of the computed waveform or queries
the current setting.
Syntax
:MATH<x>:SCALe:CENTer {<NRf>|<Voltage>|
<Current>}
:MATH<x>:SCALe:CENTer?
<x> = 1 to 8
<NRf>, <V
oltage>, and <Current> = See the DL9000
User’s Manual.
Example
:MATH1:SCALE:CENTER 1
:MATH1:SCALE:CENTER? -> :MATH1:SCALE:
CENTER 1.000E+00
:MATH<x>:SCALe:MODE
Function Sets the scaling mode or queries the current setting.
Syntax
:MATH<x>:SCALe:MODE {AUTO|MANual}
:MATH<x>:SCALe:MODE?
<x> = 1 to 8
Example
:MATH1:SCALE:MODE AUTO
:MATH1:SCALE:MODE? -> :MATH1:SCALE:
MODE AUTO
:MATH<x>:SCALe:SENSitivity
Function Sets the vertical sensitivity of the computed waveform
or queries the current setting.
Syntax
:MATH<x>:SCALe:SENSitivity {<NRf>|
<Voltage>|<Current>}
:MATH<x>:SCALe:SENSitivity?
<x> = 1 to 8
<NRf>, <V
oltage>, and <Current> = See the DL9000
User’s Manual.
Example
:MATH1:SCALE:SENSITIVITY 1
:MATH1:SCALE:SENSITIVITY? -> :MATH1:
SCALE:SENSITIVITY 1.000E+00
:MATH<x>:SELect
Function Sets the display option or queries the current setting.
Syntax
:MATH<x>:SELect {INPut|MATH|REFerence}
:MATH<x>:SELect?
<x> = 1 to 8
Example
:MATH1:SELECT INPUT
:MATH1:SELECT? -> :MATH1:SELECT INPUT
:MATH<x>:SVALue (Scale VALUE)
Function Turns ON/OFF the scale value display or queries the
current setting.
Syntax
:MATH<x>:SVALue {<Boolean>}
:MATH<x>:SVALue?
<x> = 1 to 8
Example
:MATH1:SVALUE ON
:MATH1:SVALUE? -> MATH1:SVALUE 1
:MATH<x>:THReshold<x>
Function Sets the edge detection level of the count
computation or queries the current setting.
Syntax
:MATH<x>:THReshold<x> {<NRf>|
<Voltage>|<Current>}
:MATH<x>:THReshold<x>?
<x> of MATH<x> = 1 to 8
<x> of THReshold<x> = 1 or 2
<NRf>, <Voltage>, and <Current> = See the DL9000
User’s Manual.
Example
:MATH1:THRESHOLD1 1
:MATH1:THRESHOLD1? -> :MATH1:
THRESHOLD1 1.000E+00
Description THReshold2 is valid when the operation is {RCOunt}.
:MATH<x>:UNIT?
Function Queries all settings related to the computation unit.
Syntax
:MATH<x>:UNIT?
<x> = 1 to 8
Example
:MATH1:UNIT? ->
:MATH1:UNIT:DEFINE "EU";MODE AUTO
:MATH<x>:UNIT[:DEFine]
Function Sets the computation unit or queries the current
setting.
Syntax
:MATH<x>:UNIT[:DEFine] {<String>}
:MATH<x>:UNIT[:DEFine]?
<x> = 1 to 8
<String> = Up to 4 characters
Example
:MATH1:UNIT:DEFINE "EU"
:MATH1:UNIT:DEFINE? -> :MATH1:UNIT:
DEFINE "EU"