EasyManuals Logo

YOKOGAWA DL9040 User Manual

YOKOGAWA DL9040
256 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #154 background imageLoading...
Page #154 background image
5-120
IM 701310-17E
:MATH<x>:POSition
Function Sets the vertical position of the computed waveform
or queries the current setting.
Syntax
:MATH<x>:POSition {<NRf>}
:MATH<x>:POSition?
<x> = 1 to 8
<NRf> = –4 to 4 (div)
Example
:MATH1:POSITION 0
:MATH1:POSITION? -> :MATH1:
POSITION 0.000E+00
:MATH<x>:SCALe?
Function Queries all settings related to scaling.
Syntax
:MATH<x>:SCALe?
<x> = 1 to 8
Example
:MATH1:SCALE? -> :MATH1:SCALE:
CENTER 1.000E+00;MODE AUTO;
SENSITIVITY 1.000E+00
:MATH<x>:SCALe:CENTer
Function Sets the offset of the computed waveform or queries
the current setting.
Syntax
:MATH<x>:SCALe:CENTer {<NRf>|<Voltage>|
<Current>}
:MATH<x>:SCALe:CENTer?
<x> = 1 to 8
<NRf>, <V
oltage>, and <Current> = See the DL9000
User’s Manual.
Example
:MATH1:SCALE:CENTER 1
:MATH1:SCALE:CENTER? -> :MATH1:SCALE:
CENTER 1.000E+00
:MATH<x>:SCALe:MODE
Function Sets the scaling mode or queries the current setting.
Syntax
:MATH<x>:SCALe:MODE {AUTO|MANual}
:MATH<x>:SCALe:MODE?
<x> = 1 to 8
Example
:MATH1:SCALE:MODE AUTO
:MATH1:SCALE:MODE? -> :MATH1:SCALE:
MODE AUTO
:MATH<x>:SCALe:SENSitivity
Function Sets the vertical sensitivity of the computed waveform
or queries the current setting.
Syntax
:MATH<x>:SCALe:SENSitivity {<NRf>|
<Voltage>|<Current>}
:MATH<x>:SCALe:SENSitivity?
<x> = 1 to 8
<NRf>, <V
oltage>, and <Current> = See the DL9000
User’s Manual.
Example
:MATH1:SCALE:SENSITIVITY 1
:MATH1:SCALE:SENSITIVITY? -> :MATH1:
SCALE:SENSITIVITY 1.000E+00
:MATH<x>:SELect
Function Sets the display option or queries the current setting.
Syntax
:MATH<x>:SELect {INPut|MATH|REFerence}
:MATH<x>:SELect?
<x> = 1 to 8
Example
:MATH1:SELECT INPUT
:MATH1:SELECT? -> :MATH1:SELECT INPUT
:MATH<x>:SVALue (Scale VALUE)
Function Turns ON/OFF the scale value display or queries the
current setting.
Syntax
:MATH<x>:SVALue {<Boolean>}
:MATH<x>:SVALue?
<x> = 1 to 8
Example
:MATH1:SVALUE ON
:MATH1:SVALUE? -> MATH1:SVALUE 1
:MATH<x>:THReshold<x>
Function Sets the edge detection level of the count
computation or queries the current setting.
Syntax
:MATH<x>:THReshold<x> {<NRf>|
<Voltage>|<Current>}
:MATH<x>:THReshold<x>?
<x> of MATH<x> = 1 to 8
<x> of THReshold<x> = 1 or 2
<NRf>, <Voltage>, and <Current> = See the DL9000
User’s Manual.
Example
:MATH1:THRESHOLD1 1
:MATH1:THRESHOLD1? -> :MATH1:
THRESHOLD1 1.000E+00
Description THReshold2 is valid when the operation is {RCOunt}.
:MATH<x>:UNIT?
Function Queries all settings related to the computation unit.
Syntax
:MATH<x>:UNIT?
<x> = 1 to 8
Example
:MATH1:UNIT? ->
:MATH1:UNIT:DEFINE "EU";MODE AUTO
:MATH<x>:UNIT[:DEFine]
Function Sets the computation unit or queries the current
setting.
Syntax
:MATH<x>:UNIT[:DEFine] {<String>}
:MATH<x>:UNIT[:DEFine]?
<x> = 1 to 8
<String> = Up to 4 characters
Example
:MATH1:UNIT:DEFINE "EU"
:MATH1:UNIT:DEFINE? -> :MATH1:UNIT:
DEFINE "EU"
5.17 MATH Group

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA DL9040 and is the answer not in the manual?

YOKOGAWA DL9040 Specifications

General IconGeneral
BrandYOKOGAWA
ModelDL9040
CategoryTest Equipment
LanguageEnglish

Related product manuals