5-162
IM 701310-17E
:TRIGger:EINTerval:EVENt<x>:ESTate:
POLarity
Function Sets the polarity of the edge/state trigger or queries
the current setting.
Syntax
:TRIGger:EINTerval:EVENt<x>:ESTate:
POLarity {ENTer|EXIT|FALL|RISE}
:TRIGger:EINTerval:EVENt<x>:ESTate:
POLarity?
<x> = 1 or 2
Example
:TRIGGER:EINTERVAL:EVENT1:ESTATE:
POLARITY ENTER
:TRIGGER:EINTERVAL:EVENT1:ESTATE:
POLARITY? -> :TRIGGER:EINTERVAL:
EVENT1:ESTATE:POLARITY ENTER
Description This command is valid when
:TRIGger:
EINTerval:EVENt<x>:TYPE EDGE
and
:
TRIGger:EINTerval:EVENt<x>:
ESTate:SOURce LINE
.
{ENT
er|EXIT} is valid when
:TRIGger:
EINTerval:EVENt<x>:TYPE EDGE|EQUalify
and
:TRIGger:SOURce:CHANnel<x>:WINDow
ON
. {FALL|RISE} is valid when
:TRIGger:
EINTerval:EVENt<x>:TYPE EDGE|EQUalify
and
:TRIGger:SOURce:CHANnel<x>:WINDow
OFF
.
{ENTer|EXIT} is valid when
:TRIGger:
EINTerval:EVENt<x>:TYPE STATe
.
:TRIGger:EINTerval:EVENt<x>:ESTate:
SOURce
Function Sets the trigger source of the edge/state trigger or
queries the current setting.
Syntax
:TRIGger:EINTerval:EVENt<x>:ESTate:
SOURce {<NRf>|EXTernal|LINE}
:TRIGger:EINTerval:EVENt<x>:ESTate:
SOURce?
<x> = 1 or 2
<NRf> = 1 to 4
Example
:TRIGGER:EINTERVAL:EVENT1:ESTATE:
SOURCE 1
:TRIGGER:EINTERVAL:EVENT1:ESTATE:
SOURCE? -> :TRIGGER:EINTERVAL:EVENT1:
ESTATE:SOURCE 1
Description This command is valid when
:TRIGger:
EINTerval:EVENt<x>:TYPE EDGE|EQUalify
.
{<NRf>|EXT
ernal|LINE} is valid when
:TRIGger:
EINTerval:EVENt<x>:TYPE EDGE
.
{<NRf>|EXTernal} is valid when
:TRIGger:
EINTerval:EVENt<x>:
TYPE EQUalify
.
:TRIGger:EINTerval:EVENt<x>:I2CBus?
Function Queries all settings related to the I
2
C bus trigger of
the event.
Syntax
:TRIGger:EINTerval:EVENt<x>:I2CBus?
<x> = 1 or 2
Example
:TRIGGER:EINTERVAL:EVENT1:I2CBUS? -> :
TRIGGER:EINTERVAL:EVENT1:I2CBUS:ADATA:
BIT10ADDRESS:PATTERN "10111011111";:
TRIGGER:EINTERVAL:EVENT1:I2CBUS:ADATA:
BIT7ADDRESS:PATTERN "11011110";:
TRIGGER:EINTERVAL:EVENT1:I2CBUS:ADATA:
BIT7APSUB:ADDRESS:PATTERN "10101011";:
TRIGGER:EINTERVAL:EVENT1:I2CBUS:ADATA:
BIT7APSUB:SADDRESS:PATTERN "10101011";:
TRIGGER:EINTERVAL:EVENT1:I2CBUS:ADATA:
TYPE BIT10ADDRESS;:TRIGGER:EINTERVAL:
EVENT1:I2CBUS:CLOCK:SOURCE 1;:TRIGGER:
EINTERVAL:EVENT1:I2CBUS:DATA:BYTE 1;
CONDITION TRUE;DPOSITION 1;MODE 1;
PATTERN1 "10101011";
PATTERN2 "10101010";
PATTERN3 "10101111";
PATTERN4 "10101011";PMODE DONTCARE;
SOURCE 1;:TRIGGER:EINTERVAL:EVENT1:
I2CBUS:GCALL:BIT7MADDRESS:
PATTERN "1010101";:TRIGGER:EINTERVAL:
EVENT1:I2CBUS:GCALL:
SBYTE BIT7MADDRESS;:TRIGGER:EINTERVAL:
EVENT1:I2CBUS:MODE ADATA;NAIGNORE:
HSMODE 1;RACCESS 1;SBYTE 1;:TRIGGER:
EINTERVAL:EVENT1:I2CBUS:SBHSMODE:
TYPE HSMODE
:TRIGger:EINTerval:EVENt<x>:I2CBus:
ADATa?
Function Queries all settings related to the address of the I
2
C
bus trigger.
Syntax
:TRIGger:EINTerval:EVENt<x>:I2CBus:
ADATa?
<x> = 1 or 2
Example
:TRIGGER:EINTERVAL:EVENT1:I2CBUS:ADATA?
-> :TRIGGER:EINTERVAL:EVENT1:I2CBUS:
ADATA:BIT10ADDRESS:
PATTERN "10111011111";:TRIGGER:
EINTERVAL:EVENT1:I2CBUS:ADATA:
BIT7ADDRESS:PATTERN "11011110";:
TRIGGER:EINTERVAL:EVENT1:I2CBUS:ADATA:
BIT7APSUB:ADDRESS:PATTERN "10101011";:
TRIGGER:EINTERVAL:EVENT1:I2CBUS:ADATA:
BIT7APSUB:SADDRESS:PATTERN "10101011";:
TRIGGER:EINTERVAL:EVENT1:I2CBUS:ADATA:
TYPE BIT10ADDRESS
5.29 TRIGger Group