Commands
1
2
3
4
5
6
App
Index
:ANALysis:AHIStogram<x>:MEASure:
CURSor:VPOSition<x>
Function Sets the vertical cursor position of the accumulated
histogram or queries the current setting.
Syntax
:ANALysis:AHIStogram<x>:MEASure:CURSor:
VPOSition<x> {<NRf>}
:ANALysis:AHIStogram<x>:MEASure:CURSor:
VPOSition<x>?
<x> of AHIStogram<x> = 1 or 2
<x> of VPOSition<x> = 1 or 2
<NRf> = –4 to 4 (div)
Example
:ANALYSIS:AHISTOGRAM1:MEASURE:CURSOR:
VPOSITION1 1
:ANALYSIS:AHISTOGRAM1:MEASURE:CURSOR:
VPOSITION1? -> :ANALYSIS:AHISTOGRAM1:
MEASURE:CURSOR:
VPOSITION1 1.000E+00
:ANALysis:AHIStogram<x>:MEASure:MODE
Function Sets the automated measurement mode of the
accumulated histogram or queries the current setting.
Syntax
:ANALysis:AHIStogram<x>:MEASure:MODE
{CURSor|OFF|PARameter}
:ANALysis:AHIStogram<x>:MEASure:MODE?
<x> = 1 or 2
Example
:ANALYSIS:AHISTOGRAM1:MEASURE:
MODE CURSOR
:ANALYSIS:AHISTOGRAM1:MEASURE:MODE?
-> :ANALYSIS:AHISTOGRAM1:MEASURE:
MODE CURSOR
:ANALysis:AHIStogram<x>:MEASure:
PARameter?
Function Queries all settings related to the automated
measurement of waveform parameters of the
accumulated histogram.
Syntax
:ANALysis:AHIStogram<x>:MEASure:
PARameter?
<x> = 1 or 2
Example
:ANALYSIS:AHISTOGRAM1:MEASURE:
PARAMETER? -> :ANALYSIS:AHISTOGRAM1:
MEASURE:PARAMETER:AREA1:MAXIMUM:
STATE 1;:ANALYSIS:AHISTOGRAM1:MEASURE:
PARAMETER:AREA1:MEAN:STATE 1;:ANALYSIS:
AHISTOGRAM1:MEASURE:PARAMETER:AREA1:
MEDIAN:STATE 1;:ANALYSIS:AHISTOGRAM1:
MEASURE:PARAMETER:AREA1:MINIMUM:
STATE 1;:ANALYSIS:AHISTOGRAM1:MEASURE:
PARAMETER:AREA1:PEAK:STATE 1;:ANALYSIS:
AHISTOGRAM1:MEASURE:PARAMETER:AREA1:
RMS:STATE 1;:ANALYSIS:AHISTOGRAM1:
MEASURE:PARAMETER:AREA1:SD2INTEG:
STATE 1;:ANALYSIS:AHISTOGRAM1:MEASURE:
PARAMETER:AREA1:SD3INTEG:STATE 1;:
ANALYSIS:AHISTOGRAM1:MEASURE:PARAMETER:
AREA1:SDEVIATION:STATE 1;:ANALYSIS:
AHISTOGRAM1:MEASURE:PARAMETER:AREA1:
SDINTEG:STATE 1;:ANALYSIS:AHISTOGRAM1:
MEASURE:PARAMETER:AREA2:MAXIMUM:
STATE 1;:ANALYSIS:AHISTOGRAM1:MEASURE:
PARAMETER:AREA2:MEAN:STATE 1;:ANALYSIS:
AHISTOGRAM1:MEASURE:PARAMETER:AREA2:
MEDIAN:STATE 1;:ANALYSIS:AHISTOGRAM1:
MEASURE:PARAMETER:AREA2:MINIMUM:
STATE 1;:ANALYSIS:AHISTOGRAM1:MEASURE:
PARAMETER:AREA2:PEAK:STATE 1;:ANALYSIS:
AHISTOGRAM1:MEASURE:PARAMETER:AREA2:
RMS:STATE 1;:ANALYSIS:AHISTOGRAM1:
MEASURE:PARAMETER:AREA2:SD2INTEG:
STATE 1;:ANALYSIS:AHISTOGRAM1:MEASURE:
PARAMETER:AREA2:SD3INTEG:STATE 1;:
ANALYSIS:AHISTOGRAM1:MEASURE:PARAMETER:
AREA2:SDEVIATION:STATE 1;:ANALYSIS:
AHISTOGRAM1:MEASURE:PARAMETER:AREA2:
SDINTEG:STATE 1.....
5.3 ANALysis Group

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA DL9040 and is the answer not in the manual?

YOKOGAWA DL9040 Specifications

General IconGeneral
BrandYOKOGAWA
ModelDL9040
CategoryTest Equipment
LanguageEnglish

Related product manuals