EasyManuals Logo

YOKOGAWA DL9040 User Manual

YOKOGAWA DL9040
256 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #86 background imageLoading...
Page #86 background image
5-52
IM 701310-17E
:ANALysis:FFT<x>:VERTical:SENSitivity
Function Sets the vertical sensitivity of the FFT computation or
queries the current setting.
Syntax
:ANALysis:FFT<x>:VERTical:SENSitivity
{<NRf>}
:ANALysis:FFT<x>:VERTical:SENSitivity?
<x> = 1 or 2
<NRf> = 0 to 1.000E+31 (dBV)
Example
:ANALYSIS:FFT1:VERTICAL:SENSITIVITY 1
:ANALYSIS:FFT1:VERTICAL:SENSITIVITY?
-> :ANALYSIS:FFT1:VERTICAL:
SENSITIVITY 1.000E+00
:ANALysis:FFT<x>:WINDow
Function Sets the window function of the FFT computation or
queries the current setting.
Syntax
:ANALysis:FFT<x>:WINDow {FLATtop|
HANNing|RECTangle}
:ANALysis:FFT<x>:WINDow?
<x> = 1 or 2
Example
:ANALYSIS:FFT1:WINDOW FLATTOP
:ANALYSIS:FFT1:WINDOW? -> :ANALYSIS:
FFT1:WINDOW FLATTOP
:ANALysis:TYPE<x>
Function Sets the analysis function type or queries the current
setting.
Syntax
:ANALysis:TYPE<x> {AHIStogram|FFT|
WPARameter|XY}
:ANALysis:TYPE<x>?
<x> = 1 or 2
Example
:ANALYSIS:TYPE1 AHISTOGRAM
:ANALYSIS:TYPE1? -> :ANALYSIS:
TYPE1 AHISTOGRAM
:ANALysis:VTDisplay
Function Turns ON/OFF the VT waveform display or queries
the current setting.
Syntax
:ANALysis:VTDisplay {<Boolean>}
:ANALysis:VTDisplay?
Example
:ANALYSIS:VTDISPLAY ON
:ANALYSIS:VTDISPLAY? -> :ANALYSIS:
VTDISPLAY 1
:ANALysis:WAIT<x>?
Function Waits for the completion of the automated
measurement with a timeout option.
Syntax
ANALysis:WAIT<x>? {<NRf>}
<x> = 1 or 2
<NRf> = 1 to 360000 (timeout value, in units of 10
ms)
Example
ANALYSIS:WAIT1? 100 ->:ANALYSIS:WAIT1 1
Description If the execution of the automated measurement
completes within the timeout value, 0 is returned; if
it is not complete or automated measurement is not
being executed, 1 is returned. Even if the timeout
value is set long, 0 is returned when the execution of
the automated measurement is complete.
:ANALysis:WPARameter<x>?
Function Queries all settings related to the waveform
parameter measurement function.
Syntax
:ANALysis:WPARameter<x>?
<x> = 1 or 2
Example
:ANALYSIS:WPARAMETER1? -> :ANALYSIS:
WPARAMETER1:CALCULATION 1;HISTOGRAM:
MEASURE:MODE CURSOR;CURSOR:C1:
POSITION 1.000E+00;STATE 1;:ANALYSIS:
WPARAMETER1:HISTOGRAM:MEASURE:CURSOR:
C2:POSITION 2.000E+00;STATE 1;:
ANALYSIS:WPARAMETER1:HISTOGRAM:MEASURE:
CURSOR:DC:STATE 1;:ANALYSIS:
WPARAMETER1:HISTOGRAM:MEASURE:CURSOR:
LINKAGE 1;:ANALYSIS:WPARAMETER1:
HISTOGRAM:MEASURE:PARAMETER:MEAN:
STATE 1;:ANALYSIS:WPARAMETER1:
HISTOGRAM:MEASURE:PARAMETER:PEAK:
STATE 1;:ANALYSIS:WPARAMETER1:
HISTOGRAM:MEASURE:PARAMETER:SD3INTEG:
STATE 1;:ANALYSIS:WPARAMETER1:
HISTOGRAM:MEASURE:PARAMETER:SDEVIATION:
STATE 1;:ANALYSIS:WPARAMETER1:
HISTOGRAM:MEASURE:PARAMETER:SDINTEG:
STATE 1;:ANALYSIS:WPARAMETER1:LIST:
MODE STATISTIC;SCROLL HORIZONTAL;:
ANALYSIS:WPARAMETER1:MODE HISTOGRAM;
TREND:HRANGE MAIN;HSPAN 1;MEASURE:
CURSOR:C1:POSITION 1.000E+00;STATE 1;:
ANALYSIS:WPARAMETER1:TREND:MEASURE:
CURSOR:C2:POSITION 2.000E+00;STATE 1;:
ANALYSIS:WPARAMETER1:TREND:MEASURE:
CURSOR:DC:STATE 1;:ANALYSIS:
WPARAMETER1:TREND:MEASURE:MODE CURSOR;:
ANALYSIS:WPARAMETER1:TREND:
VERTICAL 2.000E+00,1.000E+00
5.3 ANALysis Group

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA DL9040 and is the answer not in the manual?

YOKOGAWA DL9040 Specifications

General IconGeneral
BrandYOKOGAWA
ModelDL9040
CategoryTest Equipment
LanguageEnglish

Related product manuals