5-53
IM 701310-17E
Commands
1
2
3
4
5
6
App
Index
:ANALysis:WPARameter<x>:CALCulation
Function Sets the calculation items of the automated
measurement of waveform parameters or queries the
current setting.
Syntax
:ANALysis:WPARameter<x>:CALCulation
{<NRf>}
:ANALysis:WPARameter<x>:CALCulation?
<x> = 1 or 2
<NRf> = 1 to 4
Example
:ANALYSIS:WPARAMETER1:CALCULATION 1
:ANALYSIS:WPARAMETER1:CALCULATION?
-> :ANALYSIS:WPARAMETER1:CALCULATION 1
:ANALysis:WPARameter<x>:HISTogram?
Function Queries all settings related to the histogram display
of the waveform parameter measurement.
Syntax
:ANALysis:WPARameter<x>:HISTogram?
<x> = 1 or 2
Example
:ANALYSIS:WPARAMETER1:HISTOGRAM?
-> :ANALYSIS:WPARAMETER1:HISTOGRAM:
MEASURE:MODE CURSOR;CURSOR:C1:
POSITION 1.000E+00;STATE 1;
:ANALYSIS:WPARAMETER1:HISTOGRAM:
MEASURE:CURSOR:C2:
POSITION 2.000E+00;STATE 1;
:ANALYSIS:WPARAMETER1:HISTOGRAM:
MEASURE:CURSOR:DC:STATE 1;
:ANALYSIS:WPARAMETER1:HISTOGRAM:
MEASURE:CURSOR:LINKAGE 1;
:ANALYSIS:WPARAMETER1:HISTOGRAM:
MEASURE:PARAMETER:MEAN:STATE 1;
:ANALYSIS:WPARAMETER1:HISTOGRAM:
MEASURE:PARAMETER:PEAK:STATE 1;
:ANALYSIS:WPARAMETER1:HISTOGRAM:
MEASURE:PARAMETER:SD3INTEG:STATE 1;
:ANALYSIS:WPARAMETER1:HISTOGRAM:
MEASURE:PARAMETER:SDEVIATION:STATE 1;
:ANALYSIS:WPARAMETER1:HISTOGRAM:
MEASURE:PARAMETER:SDINTEG:STATE 1
:ANALysis:WPARameter<x>:HISTogram:
MEASure?
Function Queries all settings related to the automated
measurement of the histogram display of the
waveform parameter measurement.
Syntax
:ANALysis:WPARameter<x>:HISTogram:
MEASure?
<x> = 1 or 2
Example
:ANALYSIS:WPARAMETER1:HISTOGRAM:
MEASURE? -> :ANALYSIS:WPARAMETER1:
HISTOGRAM:
MEASURE:MODE CURSOR;CURSOR:C1:
POSITION 1.000E+00;STATE 1;
:ANALYSIS:WPARAMETER1:HISTOGRAM:
MEASURE:CURSOR:C2:POSITION 2.000E+00;
STATE 1;
:ANALYSIS:WPARAMETER1:HISTOGRAM:
MEASURE:CURSOR:DC:STATE 1;
:ANALYSIS:WPARAMETER1:HISTOGRAM:
MEASURE:CURSOR:LINKAGE 1;
:ANALYSIS:WPARAMETER1:HISTOGRAM:
MEASURE:PARAMETER:MEAN:STATE 1;
:ANALYSIS:WPARAMETER1:HISTOGRAM:
MEASURE:PARAMETER:PEAK:STATE 1;
:ANALYSIS:WPARAMETER1:HISTOGRAM:
MEASURE:PARAMETER:SD3INTEG:STATE 1;
:ANALYSIS:WPARAMETER1:HISTOGRAM:
MEASURE:PARAMETER:SDEVIATION:STATE 1;
:ANALYSIS:WPARAMETER1:HISTOGRAM:
MEASURE:PARAMETER:SDINTEG:STATE 1
:ANALysis:WPARameter<x>:HISTogram:
MEASure:CURSor?
Function Queries all settings related to the cursor
measurement on the histogram of the waveform
parameter measurement.
Syntax
:ANALysis:WPARameter<x>:HISTogram:
MEASure:CURSor?
<x> = 1 or 2
Example
:ANALYSIS:WPARAMETER1:HISTOGRAM:
MEASURE:CURSOR? -> :ANALYSIS:
WPARAMETER1:HISTOGRAM:
MEASURE:CURSOR:C1:POSITION 1.000E+00;
STATE 1;
:ANALYSIS:WPARAMETER1:HISTOGRAM:
MEASURE:CURSOR:C2:POSITION 2.000E+00;
STATE 1;
:ANALYSIS:WPARAMETER1:HISTOGRAM:
MEASURE:CURSOR:DC:STATE 1;
:ANALYSIS:WPARAMETER1:HISTOGRAM:
MEASURE:CURSOR:LINKAGE 1
5.3 ANALysis Group

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA DL9040 and is the answer not in the manual?

YOKOGAWA DL9040 Specifications

General IconGeneral
BrandYOKOGAWA
ModelDL9040
CategoryTest Equipment
LanguageEnglish

Related product manuals