5-122
IM 710105-17E
:LOGic:GROup:STATe:THReshold
Function Sets or queries the detection level of the reference
clock waveform of the logic-input state display.
Syntax
:LOGic:GROup:STATe:THReshold {<NRf>}
:LOGic:GROup:STATe:THReshold?
<NRf> = See the DLM2000 Features Guide for
this information.
Example
:LOGIC:GROUP:STATE:THRESHOLD 1
:LOGIC:GROUP:STATE:THRESHOLD?
-> :LOGIC:GROUP:STATE:THRESHOLD 1.000
E+00
:LOGic:POSition
Function Sets or queries the vertical position of the logic
signal.
Syntax
:LOGic:POSition {<NRf>}
:LOGic:POSition?
<NRf>=−7to39
Example
:LOGIC:POSITION 0
:LOGIC:POSITION?
-> :LOGIC:POSITION 0
:LOGic:SIZE
Function Sets or queries the display size of the logic signal.
Syntax
:LOGic:SIZE {LARGe|MIDium|SMALl}
:LOGic:SIZE?
Example
:LOGIC:SIZE LARGE
:LOGIC:SIZE?
-> :LOGIC:SIZE LARGE
5.18 LOGic Group