5-231
IM 710105-17E
Commands
1
2
3
4
5
6
App
Index
5.32 TIMebase Group
:TIMebase?
Function Queries all time base settings.
Syntax
:TIMebase?
:TIMebase:SRATe? (Sample RATE)
Function Queries the sample rate.
Syntax
:TIMebase:SRATe?
Example
:TIMEBASE:SRATE?
-> :TIMEBASE:SRATE 12.50E+06
:TIMebase:TDIV
Function Sets or queries the Time/div value.
Syntax
:TIMebase:TDIV {<Time>}
:TIMebase:TDIV?
<Time> = 500 ps to 500 s
Example
:TIMEBASE:TDIV 2NS
:TIMEBASE:TDIV?
-> :TIMEBASE:TDIV2.000E−06