EasyManuals Logo
Home>YOKOGAWA>Test Equipment>DLM2000 Series

YOKOGAWA DLM2000 Series User Manual

YOKOGAWA DLM2000 Series
340 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #158 background imageLoading...
Page #158 background image
5-125
IM 710105-17E
Commands
1
2
3
4
5
6
App
Index
:MATH<x>:RCOunt:THReshold<y>
Function Sets or queries the threshold level for rotary-count
computation.
Syntax
:MATH<x>:RCOunt:THReshold<y> {<NRf>|
<Voltage>|<Current>}
:MATH<x>:RCOunt:THReshold<y>?
<x> = 1 or 2 (1 only for 2-channel models)
<y> = 1 or 2
<Voltage>, <Current> = See the DLM2000
Features Guide for this information.
Example
:MATH1:RCOUNT:THRESHOLD1 1
:MATH1:RCOUNT:THRESHOLD1?
-> :MATH1:RCOUNT:THRESHOLD1 1.000E+00
:MATH<x>:SCALe?
Function Queries all scaling settings.
Syntax
:MATH<x>:SCALe?
<x> = 1 or 2 (1 only for 2-channel models)
:MATH<x>:SCALe:CENTer
Function Sets or queries a center value for manual scaling.
Syntax
:MATH<x>:SCALe:CENTer {<NRf>}
:MATH<x>:SCALe:CENTer?
<x> = 1 or 2 (1 only for 2-channel models)
<NRf>=−1.0000E+31to1.0000E+31
Example
:MATH1:SCALE:CENTER 1
:MATH1:SCALE:CENTER?
-> :MATH1:SCALE:CENTER 1.00000E+00
:MATH<x>:SCALe:MODE
Function Sets or queries a scaling mode.
Syntax
:MATH<x>:SCALe:MODE {AUTO|MANual}
:MATH<x>:SCALe:MODE?
<x> = 1 or 2 (1 only for 2-channel models)
Example
:MATH1:SCALE:MODE AUTO
:MATH1:SCALE:MODE?
-> :MATH1:SCALE:MODE AUTO
:MATH<x>:SCALe:SENSitivity
Function Sets or queries a span from the center value for
manual scaling.
Syntax
:MATH<x>:SCALe:SENSitivity {<NRf>}
:MATH<x>:SCALe:SENSitivity?
<x> = 1 or 2 (1 only for 2-channel models)
<NRf>=−1.0000E+31to1.0000E+31
Example
:MATH1:SCALE:SENSITIVITY 10
:MATH1:SCALE:SENSITIVITY?
-> :MATH1:SCALE:SENSITIVITY 10.0000E+
00
:MATH<x>:UNIT?
Function Queries all computation unit settings.
Syntax
:MATH<x>:UNIT?
<x> = 1 or 2 (1 only for 2-channel models)
:MATH<x>:UNIT[:DEFine]
Function Sets or queries a unit of computation.
Syntax
:MATH<x>:UNIT[:DEFine] {<String>}
:MATH<x>:UNIT[:DEFine]?
<x> = 1 or 2 (1 only for 2-channel models)
<String> = Up to four characters
Example
:MATH1:UNIT:DEFINE "EU"
:MATH1:UNIT:DEFINE?
-> :MATH1:UNIT:DEFINE "EU"
Description Units are applied to scale values. They never
affect the computation result.
:MATH<x>:UNIT:MODE
Function Sets or queries whether a unit of computation will
be attached automatically or manually.
Syntax
:MATH<x>:UNIT:MODE {AUTO|USERdefine}
:MATH<x>:UNIT:MODE?
<x> = 1 or 2 (1 only for 2-channel models)
Example
:MATH1:UNIT:MODE AUTO
:MATH1:UNIT:MODE?
-> :MATH1:UNIT:MODE AUTO
:MATH<x>:USERdefine?
Function Queries all user-defined computation settings.
Syntax
:MATH<x>:USERdefine?
<x> = 1 or 2 (1 only for 2-channel models)
:MATH<x>:USERdefine:AVERage?
Function Queries all averaging settings for user-defined
computation.
Syntax
:MATH<x>:USERdefine:AVERage?
<x> = 1 or 2 (1 only for 2-channel models)
Description The command affects both MATH1 and MATH2.
5.19 MATH Group

Table of Contents

Other manuals for YOKOGAWA DLM2000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA DLM2000 Series and is the answer not in the manual?

YOKOGAWA DLM2000 Series Specifications

General IconGeneral
BrandYOKOGAWA
ModelDLM2000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals